Comparison of Pulmonary Compliance Measured by Multiple Linear Regression Versus Traditional Methods in Mechanically Ventilated Patients

Abstract

Background: Pulmonary compliance measurement is a critical component of monitoring mechanically ventilated patients with respiratory failure. The traditional method calculates compliance by dividing delivered tidal volume by the resultant airway pressure (plateau pressure minus positive end-expiratory pressure [PEEP]). However, this approach requires intermittent ventilator disconnection, limiting its frequency. A novel method using multiple linear regression (MLR) analysis of continuous pressure and flow waveforms enables breath-to-breath compliance measurement without disrupting ventilation. This study compares pulmonary compliance values obtained by MLR and traditional methods.
Methods: In this clinical study, pulmonary compliance was measured and compared in consecutive mechanically ventilated patients using both traditional and MLR methods. MLR-derived compliance was obtained using the ventilator’s integrated monitoring function, while traditional compliance was calculated as tidal volume divided by (plateau pressure – PEEP).
Results: Among 200 enrolled patients, the two methods showed strong correlation (*r* = 0.9, *p* < 0.01). However, MLR-derived compliance values were consistently lower than those from the traditional method (44.74 ± 21.78 mL/cmH₂O vs. 57.95 ± 26.64 mL/cmH₂O, *p* < 0.01).
Conclusion: MLR is a reliable alternative for continuous pulmonary compliance monitoring, though its systematically lower values—likely reflecting dynamic rather than static compliance—may necessitate a correction factor. The method’s ability to provide breath-to-breath measurements offers significant clinical advantages over traditional intermittent assessments.

[1] Slutsky AS, Ranieri VM. Ventilator-induced lung injury. N Engl J Med 2013;369:2126-36.
[2] Tobin MJ. Advances in mechanical ventilation. N Engl J Med. 2001;344(26):1986-96.
[3] Harris RS. Pressure-volume curves of the respiratory system. Respir Care. 2005;50(1):78-98.
[4] Hess DR. Respiratory mechanics in mechanically ventilated patients. Respir Care. 2014;59(11):1773-94.
[5] Hezarjaribi N, Dutta R, Tao X, Murdoch GK, Mazrouee S, Mortazavi B, et al. Monitoring Lung Mechanics during Mechanical Ventilation using Machine Learning Algorithms, 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Honolulu, HI, USA, 2018; 1160-1163.
[6] Yi B, Wang C, Gao Y, Gong Y, Luo L, Yao R, et al. Continuous estimation of respiratory system compliance and airway resistance using multiple linear regression. BMC Pulm Med. 2024;24(1):106.
[7] Chen ZL, Yan YZ, Yu HY, Wang QB, Wang W, Zhong M. Influence of compliance and resistance of the test lung on the accuracy of the tidal volume delivered by the ventilator. BMC Pulm Med. 2024;24(1):498.
[8] Polak AG. Analysis of multiple linear regression algorithms used for respiratory mechanics monitoring during artificial ventilation. Comput Methods Programs Biomed. 2011;101(2):126-34.
[9] Nikischin W, Gerhardt T, Everett R, Bancalari E. A new method to analyze lung compliance when pressure-volume relationship is nonlinear. Am J Respir Crit Care Med. 1998;158(4):1052-60.
[10] Chen Y, Yuan Y, Chang Q, Zhang H, Li F, Chen Z. Continuous estimation of respiratory system compliance and airway resistance during pressure-controlled ventilation without end-inspiration occlusion. BMC Pulm Med. 2024; 24(1):249.
[11] Rousselot JM, Peslin R, Duvivier C. Evaluation of the multiple linear regression method to monitor respiratory mechanics in ventilated neonates and young children. Pediatr Pulmonol. 1992; 13(3):161-8.
[12] Kulkarni A, Sheela N. Methods to Evaluate Airway Resistance and Lung Compliance During Mechanical Ventilation: A Comparative Study. Int J Innov Sci Res Technol. 2020; 5(7):947-951.
[13] Acute Respiratory Distress Syndrome Network; Brower RG, Matthay MA, Morris A, Schoenfeld D, Thompson BT, Wheeler A. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med. 2000; 342(18):1301-8.
[14] Gutierrez, G. A novel method to calculate compliance and airway resistance in ventilated patients Intensive Care Med Exp. 2022; 10(1):55.
[15] Bradbrook CA, Clark L, Dugdale AH, Burford J, Mosing M. Measurement of respiratory system compliance and respiratory system resistance in healthy dogs undergoing general anaesthesia for elective orthopaedic procedures. Vet Anaesth Analg. 2013;40(4):382-9.
[16] Officer TM, Pellegrino R, Brusasco V, Rodarte JR. Measurement of pulmonary resistance and dynamic compliance with airway obstruction. J Appl Physiol (1985). 1998;85(5):1982-8.
[17] Stenqvist O., Practical assessment of respiratory mechanics, Br J Anaesth. 2003;91(1):92-105.
[18] Kamp L, Hunnekens B, Wouw N, Oomen T. An Estimation Perspective on Breathing Effort Disturbances in Mechanical Ventilation. IFAC-PapersOnLine. 2023; 56(2): 8215-8220.
Files
IssueArticle in Press QRcode
SectionResearch Article(s)
Keywords
Pulmonary compliance Mechanical ventilation Respiratory mechanics Respiratory failure Multiple linear regression

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
Habibzadeh MR, Saghaei M, Shojaei M. Comparison of Pulmonary Compliance Measured by Multiple Linear Regression Versus Traditional Methods in Mechanically Ventilated Patients. Arch Anesth & Crit Care. 2025;.