Correlation of Ultrasound Guided Inferior Vena Cava Collapsibility Index with Central Venous Pressure to Assess the Volume Status in Postoperative Intensive Care Unit Patients: A Prospective Observational Study

Abstract

Background: Accurate assessment of intravascular volume status is a vital aspect of management of intensive care unit (ICU) patients. Inferior vena cava (IVC) diameter and IVC collapsibility index have surfaced as promising methods to accurately predict hypovolemia. But no such study has evaluated it’s their utility in postoperative patients.
Methods: The study aimed to assess the correlation between ultrasound guided IVC collapsibility index (IVC CI) and CVP for volume status in intensive care unit patients. Hundred spontaneously breathing patients receiving postoperative care in our surgical ICU between November 2019 to march 2021were enrolled into the study. Maximum IVC diameter (IVCdmax) at end-expiration, minimum IVC diameter at end-inspiration (IVCdmin) and IVC CI were measured. Simultaneous CVP recordings were obtained.
Results: A positive correlation was noted between IVC maximum diameter and CVP (p = <0.001) and between IVC Minimum Diameter and CVP. (p = <0.001) A negative correlation between IVC CI and CVP was seen (p = <0.001). Mean IVC CI was highest in the hypovolemic group. The area under the ROC curve (AUROC) for IVC CI predicting hypovolemia was 0.943 (95% CI: 0.9 - 0.986), thus demonstrating excellent diagnostic performance. At a cut off of ≥58.416%, IVC CI predicts hypovolemia with a sensitivity of 93.8%, and a specificity of 84%.
Conclusion: IVC CI can be used to guide fluid therapy due to its excellent diagnostic accuracy in predicting hypovolemia in postoperative patients in ICU.

[1] Sessler DI, Meyhoff CS, Zimmerman NM, Mao G, Leslie K, Vásquez SM, et al. Period- dependent associations between hypotension during and for four days after noncardiac surgery and a composite of myocardial infarction and death: a substudy of the POISE-2 trial. Anesthesiology. 2018; 128:317-27.
[2] Rivers E, Nguyen B, Havstad S, Ressler J, Muzzin A, Knoblich B, et al. Early goal-directed therapy in the treatment of severe sepsis and septic shock. N Engl J Med. 2001; 345:1368-77.
[3] Kalantari K, Chang J, Ronco C, Rosner M. Assessment of intravascular volume status and volume responsiveness in critically ill patients. Kidney Int. 2013; 83:1017-1028.
[4] Marik PE, Baram M, Vahid B. Does central venous pressure predict fluid responsiveness? A systematic review of the literature and the tale of seven mares. Chest. 2008; 134:172-8.
[5] Marik PE, Baram M. Non-invasive hemodynamic monitoring in the intensive care unit. Crit Care Clin. 2007; 23: 383–400.
[6] Marik PE, Flemmer M, Harrison W. The risk of catheter-related bloodstream infection with femoral venous catheters as compared to subclavian and internal jugular venous catheters: a systematic review of the literature and meta-analysis. Crit care med. 2012; 40:2479-85.
[7] Nagdev AD, Merchant RC, Tirado-Gonzalez A, Sisson CA, Murphy MC. Emergency department bedside ultrasonographic measurement of the caval index for noninvasive determination of low central venous pressure. Ann Emerg Med. 2010; 55:290-5.
[8] Fields JM, Lee PA, Jenq KY, Mark DG, Panebianco NL, Dean AJ. The interrater reliability of inferior vena cava ultrasound by bedside clinician sonographers in emergency department patients. Acad Emerg Med. 2011; 18:98-101.
[9] Ilyas A, Ishtiaq W, Assad S, Ghazanfar H, Mansoor S, Haris M, et al. Correlation of IVC Diameter and Collapsibility Index with Central Venous Pressure in the Assessment of Intravascular Volume in Critically Ill Patients. Cureus. 9: e1025.
[10] Pasquero P, Albani S, Sitia E, Castagno F, D’Antico S, Porta M. Inferior vena cava diameters and collapsibility index changes reveal early volume depletion in a healthy donor model. Crit Ultrasound J. 2012; 18:A29.
[11] Thanakitcharu P, Charoenwut M, Siriwiwatanakul N. Inferior vena cava diameter and collapsibility index: a practical non-invasive evaluation of intravascular fluid volume in critically-ill patients. J Med Assoc Thai. 2013; 96:S14-22.
[12] Stawicki SP, Braslow BM, Panebianco NL, Kirkpatrick JN, Gracias VH, Hayden GE, et al. Intensivist use of hand-carried ultrasonography to measure IVC collapsibility in estimating intravascular volume status: correlations with CVP. J Am Coll Surg. 2009; 209:55-61.
[13] Nagi AI, Shafik AM, Fatah AMA, Selima WZ, Hefny AF, et al. Inferior vena cava collapsibility index as a predictor of fluid responsiveness in sepsis-related acute circulatory failure. Ain-Shams J Anesthesiol. 2021; 13:75.
[14] Szabó M, Bozó A, Darvas K, Horváth A, Iványi Z. Role of inferior vena cava collapsibility index in the prediction of hypotension associated with general anesthesia: an observational study. BMC Anesthesiol. 2019; 19:139.
[15] Airapetian N, Maizel J, Alyamani O, Mahjoub Y, Lorne E, Levrard M, et al. Does inferior vena cava respiratory variability predict fluid responsiveness in spontaneously breathing patients? Crit Care. 2015; 19:400.
[16] Beigel R, Cercek B, Luo H, Siegel RJ. Noninvasive evaluation of right atrial pressure. J Am Soc Echocardiogr. 2013 Sep;26(9):1033-42.
Files
IssueVol 10 No 1 (2024): Winter QRcode
SectionResearch Article(s)
DOI https://doi.org/10.18502/aacc.v10i1.14769
Keywords
Collapsibility index Intensive care unit Ultrasonography Inferior vena cava Central venous pressure

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
Sree Surendran A, Kaur R, Yadav R. Correlation of Ultrasound Guided Inferior Vena Cava Collapsibility Index with Central Venous Pressure to Assess the Volume Status in Postoperative Intensive Care Unit Patients: A Prospective Observational Study. Arch Anesth & Crit Care. 2023;10(1):30-35.